Gas chromatography–mass spectrometry ( GC–MS ) is an analytical method that combines the features of gas-chromatography and mass spectrometry to identify different substances within a test sample. Applications of GC–MS include drug detection, fire investigation, environmental analysis, explosives investigation, food and flavor analysis, and identification of unknown samples, including that of material samples obtained from planet Mars during probe missions as early as the 1970s. GC–MS can also be used in airport security to detect substances in luggage or on human beings. Additionally, it can identify trace elements in materials that were previously thought to have disintegrated beyond identification. Like liquid chromatography–mass spectrometry , it allows analysis and detection even of tiny amounts of a substance.

#578421

102-584: GCMS may refer to: Gas chromatography–mass spectrometry , an analytical method to identify different substances within a test sample Gibson City-Melvin-Sibley Community Unit School District 5 , a K–12 public school district based in Gibson City, Illinois Global Case Management System is a software system used by Immigration, Refugees and Citizenship Canada (IRCC) to process immigration and citizenship applications. Information extracted from this software system

204-524: A category of substances is present. Although a nonspecific test could statistically suggest the identity of the substance, this could lead to false positive identification. However, the high temperatures (300°C) used in the GC–MS injection port (and oven) can result in thermal degradation of injected molecules, thus resulting in the measurement of degradation products instead of the actual molecule(s) of interest. The first on-line coupling of gas chromatography to

306-414: A certain compound. This is a fast and efficient analysis, especially if the analyst has previous information about a sample or is only looking for a few specific substances. When the amount of information collected about the ions in a given gas chromatographic peak decreases, the sensitivity of the analysis increases. So, SIM analysis allows for a smaller quantity of a compound to be detected and measured, but

408-480: A charcoal column that also used mercury. Gerhard Hesse, while a professor at the University of Marburg /Lahn decided to test the prevailing opinion among German chemists that molecules could not be separated in a moving gas stream. He set up a simple glass column filled with starch and successfully separated bromine and iodine using nitrogen as the carrier gas. He then built a system that flowed an inert gas through

510-467: A chemical formula containing the various elements which are believed to be in the compound. The isotope pattern in the spectrum, which is unique for elements that have many natural isotopes, can also be used to identify the various elements present. Once a chemical formula has been matched to the spectrum, the molecular structure and bonding can be identified, and must be consistent with the characteristics recorded by GC–MS. Typically, this identification

612-413: A constant sensitivity over long period of time. In addition, when alkali ions are not added to the flame, AFD operates like a standard FID. A catalytic combustion detector (CCD) measures combustible hydrocarbons and hydrogen. Discharge ionization detector (DID) uses a high-voltage electric discharge to produce ions. Flame photometric detector (FPD) uses a photomultiplier tube to detect spectral lines of

714-524: A current between the electrodes. The increase in current is translated and appears as a peak in a chromatogram. FIDs have low detection limits (a few picograms per second) but they are unable to generate ions from carbonyl containing carbons. FID compatible carrier gasses include helium, hydrogen, nitrogen, and argon. In FID, sometimes the stream is modified before entering the detector. A methanizer converts carbon monoxide and carbon dioxide into methane so that it can be detected. A different technology

816-438: A detector response. Nitrogen–phosphorus detector (NPD), a form of thermionic detector where nitrogen and phosphorus alter the work function on a specially coated bead and a resulting current is measured. Dry electrolytic conductivity detector (DELCD) uses an air phase and high temperature (v. Coulsen) to measure chlorinated compounds. Mass spectrometer (MS), also called GC-MS ; highly effective and sensitive, even in

918-655: A gas chromatograph and quadrupole mass spectrometer that can be used in tandem as a GC–MS. The material in the comet 67P/Churyumov–Gerasimenko was analysed by the Rosetta mission with a chiral GC–MS in 2014. Dozens of congenital metabolic diseases also known as inborn errors of metabolism (IEM) are now detectable by newborn screening tests, especially the testing using gas chromatography–mass spectrometry. GC–MS can determine compounds in urine even in minor concentration. These compounds are normally not present but appear in individuals suffering with metabolic disorders. This

1020-427: A gas switching valve system; adsorbed samples (e.g., on adsorbent tubes) are introduced using either an external (on-line or off-line) desorption apparatus such as a purge-and-trap system, or are desorbed in the injector (SPME applications). The real chromatographic analysis starts with the introduction of the sample onto the column. The development of capillary gas chromatography resulted in many practical problems with

1122-545: A glass condenser packed with silica gel and collected the eluted fractions. Courtenay S.G Phillips of Oxford University investigated separation in a charcoal column using a thermal conductivity detector. He consulted with Claesson and decided to use displacement as his separating principle. After learning about the results of James and Martin, he switched to partition chromatography. Early gas chromatography used packed columns, made of block 1–5 m long, 1–5 mm diameter, and filled with particles. The resolution of packed columns

SECTION 10

#1732780250579

1224-418: A mass spectrometer was reported in the late 1950s. An interest in coupling the methods had been suggested as early as December 1954, but conventional recording techniques had too poor temporal resolution. Fortunately, time-of-flight mass spectrometry developed around the same time allowed to measure spectra thousands times a second. The development of affordable and miniaturized computers has helped in

1326-434: A much finer degree of substance identification than either unit used separately. It is not possible to make an accurate identification of a particular molecule by gas chromatography or mass spectrometry alone. The mass spectrometry process normally requires a very pure sample while gas chromatography using a traditional detector (e.g. Flame ionization detector ) cannot differentiate between multiple molecules that happen to take

1428-404: A number of problems inherent in the use of syringes for injection. Even the best syringes claim an accuracy of only 3%, and in unskilled hands, errors are much larger. The needle may cut small pieces of rubber from the septum as it injects sample through it. These can block the needle and prevent the syringe filling the next time it is used. It may not be obvious that this has happened. A fraction of

1530-608: A part of all US airports . These systems run on a host of technologies, many of them based on GC–MS. There are only three manufacturers certified by the FAA to provide these systems, one of which is Thermo Detection (formerly Thermedics), which produces the EGIS , a GC–MS-based line of explosives detectors. The other two manufacturers are Barringer Technologies, now owned by Smith's Detection Systems, and Ion Track Instruments, part of General Electric Infrastructure Security Systems. As part of

1632-438: A prominent product ion which can be the product ion for the chosen precursor ion. The pair is called a "transition" and forms the basis for SRM. SRM is highly specific and virtually eliminates matrix background. After the molecules travel the length of the column, pass through the transfer line and enter into the mass spectrometer they are ionized by various methods with typically only one method being used at any given time. Once

1734-440: A radioactive beta particle (electron) source to measure the degree of electron capture. ECD are used for the detection of molecules containing electronegative / withdrawing elements and functional groups like halogens, carbonyl, nitriles, nitro groups, and organometalics. In this type of detector either nitrogen or 5% methane in argon is used as the mobile phase carrier gas. The carrier gas passes between two electrodes placed at

1836-436: A second phase of mass fragmentation is added, for example using a second quadrupole in a quadrupole instrument, it is called tandem MS (MS/MS). MS/MS can sometimes be used to quantitate low levels of target compounds in the presence of a high sample matrix background. The first quadrupole (Q1) is connected with a collision cell (Q2) and another quadrupole (Q3). Both quadrupoles can be used in scanning or static mode, depending on

1938-507: A small number of samples), to robotic technologies (XYZ robot vs. rotating robot – the most common), or to analysis: The column inlet (or injector) provides the means to introduce a sample into a continuous flow of carrier gas. The inlet is a piece of hardware attached to the column head. Common inlet types are: The choice of carrier gas (mobile phase) is important. Hydrogen has a range of flow rates that are comparable to helium in efficiency. However, helium may be more efficient and provide

2040-423: A small quantity of sample. This detector can be used to identify the analytes in chromatograms by their mass spectrum. Some GC-MS are connected to an NMR spectrometer which acts as a backup detector. This combination is known as GC-MS-NMR . Some GC-MS-NMR are connected to an infrared spectrophotometer which acts as a backup detector. This combination is known as GC-MS-NMR-IR. It must, however, be stressed this

2142-448: A substance can be measured, but it is often required that the sample must be measured in comparison to a sample containing the pure, suspected substance known as a reference standard . Various temperature programs can be used to make the readings more meaningful; for example to differentiate between substances that behave similarly during the GC process. Professionals working with GC analyze

SECTION 20

#1732780250579

2244-515: A temperature controlled oven. As the chemicals exit the end of the column, they are detected and identified electronically. Chromatography dates to 1903 in the work of the Russian scientist, Mikhail Semenovich Tswett , who separated plant pigments via liquid column chromatography. The invention of gas chromatography is generally attributed to Anthony T. James and Archer J.P. Martin . Their gas chromatograph used partition chromatography as

2346-488: A wide range of mass fragments. Full scan is useful in determining unknown compounds in a sample. It provides more information than SIM when it comes to confirming or resolving compounds in a sample. During instrument method development it may be common to first analyze test solutions in full scan mode to determine the retention time and the mass fragment fingerprint before moving to a SIM instrument method. In selective ion monitoring (SIM) certain ion fragments are entered into

2448-452: Is a quadrupole or the ion trap itself in an ion trap MS) where they are bombarded with free electrons emitted from a filament, not unlike the filament one would find in a standard light bulb. The electrons bombard the molecules, causing the molecule to fragment in a characteristic and reproducible way. This "hard ionization" technique results in the creation of more fragments of low mass-to-charge ratio (m/z) and few, if any, molecules approaching

2550-558: Is also commonly used in forensic toxicology to find drugs and/or poisons in biological specimens of suspects, victims, or the deceased. In drug screening, GC–MS methods frequently utilize liquid-liquid extraction as a part of sample preparation, in which target compounds are extracted from blood plasma. GC–MS is the main tool used in sports anti-doping laboratories to test athletes' urine samples for prohibited performance-enhancing drugs, for example anabolic steroids . A post–September 11 development, explosive detection systems have become

2652-464: Is also frequently determined by the detector, though the level of sensitivity needed can also play a significant role. Typically, purities of 99.995% or higher are used. The most common purity grades required by modern instruments for the majority of sensitivities are 5.0 grades, or 99.999% pure meaning that there is a total of 10 ppm of impurities in the carrier gas that could affect the results. The highest purity grades in common use are 6.0 grades, but

2754-665: Is also used to detect and measure contaminants from spoilage or adulteration which may be harmful and which is often controlled by governmental agencies, for example pesticides . Several GC–MS systems have left earth. Two were brought to Mars by the Viking program . Venera 11 and 12 and Pioneer Venus analysed the atmosphere of Venus with GC–MS. The Huygens probe of the Cassini–Huygens mission landed one GC–MS on Saturn 's largest moon, Titan . The MSL Curiosity rover's Sample analysis at Mars (SAM) instrument contains both

2856-413: Is also utilized in the identification of continuous phase component in a smart material, magnetorheological (MR) fluid . Foods and beverages contain numerous aromatic compounds , some naturally present in the raw materials and some forming during processing. GC–MS is extensively used for the analysis of these compounds which include esters , fatty acids , alcohols , aldehydes , terpenes etc. It

2958-449: Is at 394 nm. With an atomic emission detector (AED), a sample eluting from a column enters a chamber which is energized by microwaves that induce a plasma. The plasma causes the analyte sample to decompose and certain elements generate an atomic emission spectra. The atomic emission spectra is diffracted by a diffraction grating and detected by a series of photomultiplier tubes or photo diodes. Electron capture detector (ECD) uses

3060-595: Is calculated by finding the response of a known amount of analyte and a constant amount of internal standard (a chemical added to the sample at a constant concentration, with a distinct retention time to the analyte). In most modern GC-MS systems, computer software is used to draw and integrate peaks, and match MS spectra to library spectra. In general, substances that vaporize below 300 °C (and therefore are stable up to that temperature) can be measured quantitatively. The samples are also required to be salt -free; they should not contain ions . Very minute amounts of

3162-420: Is capable of performing both functions either individually or concomitantly, depending on the setup of the particular instrument. The primary goal of instrument analysis is to quantify an amount of substance. This is done by comparing the relative concentrations among the atomic masses in the generated spectrum. Two kinds of analysis are possible, comparative and original. Comparative analysis essentially compares

GCMS - Misplaced Pages Continue

3264-440: Is composed of two major building blocks: the gas chromatograph and the mass spectrometer . The gas chromatograph utilizes a capillary column whose properties regarding molecule separation depend on the column's dimensions (length, diameter, film thickness) as well as the phase properties (e.g. 5% phenyl polysiloxane). The difference in the chemical properties between different molecules in a mixture and their relative affinity for

3366-401: Is contained inside of a separation column. Today, most GC columns are fused silica capillaries with an inner diameter of 100–320 micrometres (0.0039–0.0126 in) and a length of 5–60 metres (16–197 ft). The GC column is located inside an oven where the temperature of the gas can be controlled and the effluent coming off the column is monitored by a suitable detector. A gas chromatograph

3468-423: Is different from Wikidata All article disambiguation pages All disambiguation pages Gas chromatography%E2%80%93mass spectrometry GC–MS has been regarded as a " gold standard " for forensic substance identification because it is used to perform a 100% specific test, which positively identifies the presence of a particular substance. A nonspecific test merely indicates that any of several in

3570-415: Is done automatically by programs which come with the instrument, given a list of the elements which could be present in the sample. A "full spectrum" analysis considers all the "peaks" within a spectrum. Conversely, selective ion monitoring (SIM) only monitors selected ions associated with a specific substance. This is done on the assumption that at a given retention time, a set of ions is characteristic of

3672-417: Is increasingly becoming a common way to diagnose IEM for earlier diagnosis and institution of treatment eventually leading to a better outcome. It is now possible to test a newborn for over 100 genetic metabolic disorders by a urine test at birth based on GC–MS. In combination with isotopic labeling of metabolic compounds, the GC–MS is used for determining metabolic activity . Most applications are based on

3774-552: Is increasingly used for detection of illegal narcotics, and may eventually supplant drug-sniffing dogs. A simple and selective GC–MS method for detecting marijuana usage was recently developed by the Robert Koch Institute in Germany. It involves identifying an acid metabolite of tetrahydrocannabinol (THC), the active ingredient in marijuana, in urine samples by employing derivatization in the sample preparation. GC–MS

3876-413: Is made of a narrow tube, known as the column , through which the vaporized sample passes, carried along by a continuous flow of inert or nonreactive gas. Components of the sample pass through the column at different rates, depending on their chemical and physical properties and the resulting interactions with the column lining or filling, called the stationary phase . The column is typically enclosed within

3978-472: Is often referred to as GCMS Notes. Topics referred to by the same term [REDACTED] This disambiguation page lists articles associated with the title GCMS . If an internal link led you here, you may wish to change the link to point directly to the intended article. Retrieved from " https://en.wikipedia.org/w/index.php?title=GCMS&oldid=1166406775 " Category : Disambiguation pages Hidden categories: Short description

4080-528: Is the collection of conditions in which the GC operates for a given analysis. Method development is the process of determining what conditions are adequate and/or ideal for the analysis required. Conditions which can be varied to accommodate a required analysis include inlet temperature, detector temperature, column temperature and temperature program, carrier gas and carrier gas flow rates, the column's stationary phase, diameter and length, inlet type and flow rates, sample size and injection technique. Depending on

4182-425: Is the ion trap mass spectrometer. Additionally one may find a magnetic sector mass spectrometer, however these particular instruments are expensive and bulky and not typically found in high-throughput service laboratories. Other detectors may be encountered such as time of flight (TOF), tandem quadrupoles (MS-MS) (see below), or in the case of an ion trap MS where n indicates the number mass spectrometry stages. When

GCMS - Misplaced Pages Continue

4284-449: Is the polyarc, by Activated Research Inc, that converts all compounds to methane. Alkali flame detector (AFD) or alkali flame ionization detector (AFID) has high sensitivity to nitrogen and phosphorus, similar to NPD. However, the alkaline metal ions are supplied with the hydrogen gas, rather than a bead above the flame. For this reason AFD does not suffer the "fatigue" of the NPD, but provides

4386-454: Is the process of separating compounds in a mixture by injecting a gaseous or liquid sample into a mobile phase, typically called the carrier gas, and passing the gas through a stationary phase. The mobile phase is usually an inert gas or an unreactive gas such as helium , argon , nitrogen or hydrogen . The stationary phase can be solid or liquid, although most GC systems today use a polymeric liquid stationary phase. The stationary phase

4488-412: Is two to three times more sensitive to analyte detection than TCD. The TCD relies on the thermal conductivity of matter passing around a thin wire of tungsten-rhenium with a current traveling through it. In this set up helium or nitrogen serve as the carrier gas because of their relatively high thermal conductivity which keep the filament cool and maintain uniform resistivity and electrical efficiency of

4590-507: Is very rare as most analyses needed can be concluded via purely GC-MS. Vacuum ultraviolet (VUV) represents the most recent development in gas chromatography detectors. Most chemical species absorb and have unique gas phase absorption cross sections in the approximately 120–240 nm VUV wavelength range monitored. Where absorption cross sections are known for analytes, the VUV detector is capable of absolute determination (without calibration) of

4692-400: The 1990s, carrier flow rate was controlled indirectly by controlling the carrier inlet pressure, or "column head pressure". The actual flow rate was measured at the outlet of the column or the detector with an electronic flow meter, or a bubble flow meter, and could be an involved, time consuming, and frustrating process. It was not possible to vary the pressure setting during the run, and thus

4794-534: The Finnigan Instrument Corporation along with Roger Sant, T. Z. Chou, Michael Story, Lloyd Friedman, and William Fies. In early 1968, they delivered the first prototype quadrupole GC/MS instruments to Stanford and Purdue University. When Finnigan Instrument Corporation was acquired by Thermo Instrument Systems (later Thermo Fisher Scientific ) in 1990, it was considered "the world's leading manufacturer of mass spectrometers". The GC–MS

4896-532: The National Institute of Standards (NIST-USA). Spectral library searches employ matching algorithms such as Probability Based Matching and dot-product matching that are used with methods of analysis written by many method standardization agencies. Sources of libraries include NIST, Wiley, the AAFS, and instrument manufacturers. The "hard ionization" process of electron ionization can be softened by

4998-465: The area of the peak using the mathematical function of integration , the concentration of an analyte in the original sample can be determined. Concentration can be calculated using a calibration curve created by finding the response for a series of concentrations of analyte, or by determining the relative response factor of an analyte. The relative response factor is the expected ratio of an analyte to an internal standard (or external standard ) and

5100-415: The best separation if flow rates are optimized. Helium is non-flammable and works with a greater number of detectors and older instruments. Therefore, helium is the most common carrier gas used. However, the price of helium has gone up considerably over recent years, causing an increasing number of chromatographers to switch to hydrogen gas. Historical use, rather than rational consideration, may contribute to

5202-408: The carrier gas. In a flame ionization detector (FID), electrodes are placed adjacent to a flame fueled by hydrogen / air near the exit of the column, and when carbon containing compounds exit the column they are pyrolyzed by the flame. This detector works only for organic / hydrocarbon containing compounds due to the ability of the carbons to form cations and electrons upon pyrolysis which generates

SECTION 50

#1732780250579

5304-406: The column temperature, the faster the sample moves through the column. However, the faster a sample moves through the column, the less it interacts with the stationary phase, and the less the analytes are separated. In general, the column temperature is selected to compromise between the length of the analysis and the level of separation. A method which holds the column at the same temperature for

5406-399: The column. Generally, chromatographic data is presented as a graph of detector response (y-axis) against retention time (x-axis), which is called a chromatogram. This provides a spectrum of peaks for a sample representing the analytes present in a sample eluting from the column at different times. Retention time can be used to identify analytes if the method conditions are constant. Also,

5508-427: The compounds as they are burned in a flame. Compounds eluting off the column are carried into a hydrogen fueled flame which excites specific elements in the molecules, and the excited elements (P,S, Halogens, Some Metals) emit light of specific characteristic wavelengths. The emitted light is filtered and detected by a photomultiplier tube. In particular, phosphorus emission is around 510–536 nm and sulfur emission

5610-413: The compounds by returning them to the liquid phase. The trap is then heated and the sample compounds are introduced to the GC–MS column via a volatiles interface, which is a split inlet system. P&T GC–MS is particularly suited to volatile organic compounds (VOCs) and BTEX compounds (aromatic compounds associated with petroleum). A faster alternative is the "purge-closed loop" system. In this system

5712-488: The continued preferential use of helium. Commonly used detectors are the flame ionization detector (FID) and the thermal conductivity detector (TCD). While TCDs are beneficial in that they are non-destructive, its low detection limit for most analytes inhibits widespread use. FIDs are sensitive primarily to hydrocarbons, and are more sensitive to them than TCD. FIDs cannot detect water or carbon dioxide which make them ideal for environmental organic analyte analysis. FID

5814-424: The cooling of the molecules before their ionization, resulting in mass spectra that are richer in information. In this method named cold electron ionization (cold-EI) the molecules exit the GC column, mixed with added helium make up gas and expand into vacuum through a specially designed supersonic nozzle, forming a supersonic molecular beam (SMB). Collisions with the make up gas at the expanding supersonic jet reduce

5916-427: The degree of certainty about the identity of that compound is reduced. When collecting data in the full scan mode, a target range of mass fragments is determined and put into the instrument's method. An example of a typical broad range of mass fragments to monitor would be m/z 50 to m/z 400. The determination of what range to use is largely dictated by what one anticipates being in the sample while being cognizant of

6018-411: The detector being used, for example, a DID requires helium as the carrier gas. When analyzing gas samples the carrier is also selected based on the sample's matrix, for example, when analyzing a mixture in argon, an argon carrier is preferred because the argon in the sample does not show up on the chromatogram. Safety and availability can also influence carrier selection. The purity of the carrier gas

6120-421: The detector(s) (see below) installed on the GC, there may be a number of detector conditions that can also be varied. Some GCs also include valves which can change the route of sample and carrier flow. The timing of the opening and closing of these valves can be important to method development. Typical carrier gases include helium , nitrogen , argon , and hydrogen . Which gas to use is usually determined by

6222-414: The electrons and analyte and cause a 'soft' ionization of the molecule of interest. A softer ionization fragments the molecule to a lower degree than the hard ionization of EI. One of the main benefits of using chemical ionization is that a mass fragment closely corresponding to the molecular weight of the analyte of interest is produced. In positive chemical ionization (PCI) the reagent gas interacts with

SECTION 60

#1732780250579

6324-425: The end of the column, and adjacent to the cathode (negative electrode) resides a radioactive foil such as 63Ni. The radioactive foil emits a beta particle (electron) which collides with and ionizes the carrier gas to generate more ions resulting in a current. When analyte molecules with electronegative / withdrawing elements or functional groups electrons are captured which results in a decrease in current generating

6426-432: The entire analysis is called "isothermal". Most methods, however, increase the column temperature during the analysis, the initial temperature, rate of temperature increase (the temperature "ramp"), and final temperature are called the temperature program. A temperature program allows analytes that elute early in the analysis to separate adequately, while shortening the time it takes for late-eluting analytes to pass through

6528-443: The filament. When analyte molecules elute from the column, mixed with carrier gas, the thermal conductivity decreases while there is an increase in filament temperature and resistivity resulting in fluctuations in voltage ultimately causing a detector response. Detector sensitivity is proportional to filament current while it is inversely proportional to the immediate environmental temperature of that detector as well as flow rate of

6630-546: The first gas chromatograph that consisted of a carrier gas, a column packed with silica gel, and a thermal conductivity detector. They exhibited the chromatograph at ACHEMA in Frankfurt, but nobody was interested in it. N.C. Turner with the Burrell Corporation introduced in 1943 a massive instrument that used a charcoal column and mercury vapors. Stig Claesson of Uppsala University published in 1946 his work on

6732-477: The flow was essentially constant during the analysis. The relation between flow rate and inlet pressure is calculated with Poiseuille's equation for compressible fluids . Many modern GCs, however, electronically measure the flow rate, and electronically control the carrier gas pressure to set the flow rate. Consequently, carrier pressures and flow rates can be adjusted during the run, creating pressure/flow programs similar to temperature programs. The polarity of

6834-406: The given spectrum to a spectrum library to see if its characteristics are present for some sample in the library. This is best performed by a computer because there are a myriad of visual distortions that can take place due to variations in scale. Computers can also simultaneously correlate more data (such as the retention times identified by GC), to more accurately relate certain data. Deep learning

6936-419: The identification probabilities of both known and unknown compounds, amplify isomer mass spectral effects and enable the use of isotope abundance analysis for the elucidation of elemental formulas. In chemical ionization (CI) a reagent gas, typically methane or ammonia is introduced into the mass spectrometer. Depending on the technique (positive CI or negative CI) chosen, this reagent gas will interact with

7038-543: The inert gas is bubbled through the water until the concentrations of organic compounds in the vapor phase are at equilibrium with concentrations in the aqueous phase. The gas phase is then analysed directly. The most common type of mass spectrometer (MS) associated with a gas chromatograph (GC) is the quadrupole mass spectrometer, sometimes referred to by the Hewlett-Packard (now Agilent ) trade name "Mass Selective Detector" (MSD). Another relatively common detector

7140-424: The injection technique. The technique of on-column injection, often used with packed columns, is usually not possible with capillary columns. In the injection system in the capillary gas chromatograph the amount injected should not overload the column and the width of the injected plug should be small compared to the spreading due to the chromatographic process. Failure to comply with this latter requirement will reduce

7242-431: The instrument method and only those mass fragments are detected by the mass spectrometer. The advantages of SIM are that the detection limit is lower since the instrument is only looking at a small number of fragments (e.g. three fragments) during each scan. More scans can take place each second. Since only a few mass fragments of interest are being monitored, matrix interferences are typically lower. To additionally confirm

7344-421: The internal vibrational (and rotational) energy of the analyte molecules, hence reducing the degree of fragmentation caused by the electrons during the ionization process. Cold-EI mass spectra are characterized by an abundant molecular ion while the usual fragmentation pattern is retained, thus making cold-EI mass spectra compatible with library search identification techniques. The enhanced molecular ions increase

7446-478: The likelihood of a potentially positive result, it is relatively important to be sure that the ion ratios of the various mass fragments are comparable to a known reference standard. GC–MS is becoming the tool of choice for tracking organic pollutants in the environment. The cost of GC–MS equipment has decreased significantly, and the reliability has increased at the same time, which has contributed to its increased adoption in environmental studies . GC–MS can analyze

7548-399: The linear velocity the faster the analysis, but the lower the separation between analytes. Selecting the linear velocity is therefore the same compromise between the level of separation and length of analysis as selecting the column temperature. The linear velocity will be implemented by means of the carrier gas flow rate, with regards to the inner diameter of the column. With GCs made before

7650-417: The mixture, but functional groups can play a large part in column selection. The polarity of the sample must closely match the polarity of the column stationary phase to increase resolution and separation while reducing run time. The separation and run time also depends on the film thickness (of the stationary phase), the column diameter and the column length. The column(s) in a GC are contained in an oven,

7752-484: The molecular mass unit. Hard ionization is considered by mass spectrometrists as the employ of molecular electron bombardment, whereas "soft ionization" is charge by molecular collision with an introduced gas. The molecular fragmentation pattern is dependent upon the electron energy applied to the system, typically 70 eV (electronvolts). The use of 70 eV facilitates comparison of generated spectra with library spectra using manufacturer-supplied software or software developed by

7854-425: The need for detection at very low levels in some forensic and environmental applications has driven the need for carrier gases at 7.0 grade purity and these are now commercially available. Trade names for typical purities include "Zero Grade", "Ultra-High Purity (UHP) Grade", "4.5 Grade" and "5.0 Grade". The carrier gas linear velocity affects the analysis in the same way that temperature does (see above). The higher

7956-694: The number of molecules present in the flow cell in the absence of chemical interferences. Olfactometric detector , also called GC-O, uses a human assessor to analyse the odour activity of compounds. With an odour port or a sniffing port, the quality of the odour, the intensity of the odour and the duration of the odour activity of a compound can be assessed. Other detectors include the Hall electrolytic conductivity detector (ElCD), helium ionization detector (HID), infrared detector (IRD), photo-ionization detector (PID), pulsed discharge ionization detector (PDD), and thermionic ionization detector (TID). The method

8058-403: The particles from a human body in order to help link a criminal to a crime . The analysis of fire debris using GC–MS is well established, and there is even an established American Society for Testing and Materials (ASTM) standard for fire debris analysis. GCMS/MS is especially useful here as samples often contain very complex matrices, and results used in court need to be highly accurate. GC–MS

8160-406: The pattern of peaks will be constant for a sample under constant conditions and can identify complex mixtures of analytes. However, in most modern applications, the GC is connected to a mass spectrometer or similar detector that is capable of identifying the analytes represented by the peaks. The area under a peak is proportional to the amount of analyte present in the chromatogram. By calculating

8262-611: The post-September 11 drive towards increased capability in homeland security and public health preparedness, traditional GC–MS units with transmission quadrupole mass spectrometers, as well as those with cylindrical ion trap (CIT-MS) and toroidal ion trap (T-ITMS) mass spectrometers have been modified for field portability and near real-time detection of chemical warfare agents (CWA) such as sarin, soman, and VX. These complex and large GC–MS systems have been modified and configured with resistively heated low thermal mass (LTM) gas chromatographs that reduce analysis time to less than ten percent of

8364-456: The purity of a particular substance, or separating the different components of a mixture. In preparative chromatography , GC can be used to prepare pure compounds from a mixture. Gas chromatography is also sometimes known as vapor-phase chromatography ( VPC ), or gas–liquid partition chromatography ( GLPC ). These alternative names, as well as their respective abbreviations, are frequently used in scientific literature. Gas chromatography

8466-406: The same amount of time to travel through the column ( i.e. have the same retention time), which results in two or more molecules that co-elute. Sometimes two different molecules can also have a similar pattern of ionized fragments in a mass spectrometer (mass spectrum). Combining the two processes reduces the possibility of error, as it is extremely unlikely that two different molecules will behave in

8568-425: The same way in both a gas chromatograph and a mass spectrometer. Therefore, when an identifying mass spectrum appears at a characteristic retention time in a GC–MS analysis, it typically increases certainty that the analyte of interest is in the sample. For the analysis of volatile compounds, a purge and trap (P&T) concentrator system may be used to introduce samples. The target analytes are extracted by mixing

8670-450: The sample is fragmented it will then be detected, usually by an electron multiplier , which essentially turns the ionized mass fragment into an electrical signal that is then detected. The ionization technique chosen is independent of using full scan or SIM. By far the most common and perhaps standard form of ionization is electron ionization (EI). The molecules enter into the MS (the source

8772-425: The sample is in liquid, gas, adsorbed, or solid form, and on whether a solvent matrix is present that has to be vaporized. Dissolved samples can be introduced directly onto the column via a COC injector, if the conditions are well known; if a solvent matrix has to be vaporized and partially removed, a S/SL injector is used (most common injection technique); gaseous samples (e.g., air cylinders) are usually injected using

8874-410: The sample may get trapped in the rubber, to be released during subsequent injections. This can give rise to ghost peaks in the chromatogram. There may be selective loss of the more volatile components of the sample by evaporation from the tip of the needle. The choice of column depends on the sample and the active measured. The main chemical attribute regarded when choosing a column is the polarity of

8976-452: The sample with water and purge with inert gas (e.g. Nitrogen gas ) into an airtight chamber, this is known as purging or sparging . The volatile compounds move into the headspace above the water and are drawn along a pressure gradient (caused by the introduction of the purge gas) out of the chamber. The volatile compounds are drawn along a heated line onto a 'trap'. The trap is a column of adsorbent material at ambient temperature that holds

9078-642: The separating principle, rather than adsorption chromatography . The popularity of gas chromatography quickly rose after the development of the flame ionization detector. Martin and another one of their colleagues, Richard Synge , with whom he shared the 1952 Nobel Prize in Chemistry , had noted in an earlier paper that chromatography might also be used to separate gases. Synge pursued other work while Martin continued his work with James. German physical chemist Erika Cremer in 1947 together with Austrian graduate student Fritz Prior developed what could be considered

9180-823: The separation capability of the column. As a general rule, the volume injected, V inj , and the volume of the detector cell, V det , should be about 1/10 of the volume occupied by the portion of sample containing the molecules of interest (analytes) when they exit the column. Some general requirements which a good injection technique should fulfill are that it should be possible to obtain the column's optimum separation efficiency, it should allow accurate and reproducible injections of small amounts of representative samples, it should induce no change in sample composition, it should not exhibit discrimination based on differences in boiling point, polarity, concentration or thermal/catalytic stability, and it should be applicable for trace analysis as well as for undiluted samples. However, there are

9282-512: The simplification of the use of this instrument, as well as allowed great improvements in the amount of time it takes to analyze a sample. In 1964, Electronic Associates, Inc. (EAI) , a leading U.S. supplier of analog computers, began development of a computer controlled quadrupole mass spectrometer under the direction of Robert E. Finnigan . By 1966 Finnigan and collaborator Mike Uthe's EAI division had sold over 500 quadrupole residual gas-analyzer instruments. In 1967, Finnigan left EAI to form

9384-423: The smaller mass detectors. Depending on the system, the analytes can be introduced via liquid injection, desorbed from sorbent tubes through a thermal desorption process, or with solid-phase micro extraction (SPME). GC–MS is used for the analysis of unknown organic compound mixtures. One critical use of this technology is the use of GC–MS to determine the composition of bio-oils processed from raw biomass. GC–MS

9486-448: The solute is crucial for the choice of stationary compound, which in an optimal case would have a similar polarity as the solute. Common stationary phases in open tubular columns are cyanopropylphenyl dimethyl polysiloxane, carbowax polyethyleneglycol, biscyanopropyl cyanopropylphenyl polysiloxane and diphenyl dimethyl polysiloxane. For packed columns more options are available. The choice of inlet type and injection technique depends on if

9588-401: The solvent and other possible interferences. A MS should not be set to look for mass fragments too low or else one may detect air (found as m/z 28 due to nitrogen), carbon dioxide ( m/z 44) or other possible interference. Additionally if one is to use a large scan range then sensitivity of the instrument is decreased due to performing fewer scans per second since each scan will have to detect

9690-581: The stationary phase of the column will promote separation of the molecules as the sample travels the length of the column. The molecules are retained by the column and then elute (come off) from the column at different times (called the retention time), and this allows the mass spectrometer downstream to capture, ionize, accelerate, deflect, and detect the ionized molecules separately. The mass spectrometer does this by breaking each molecule into ionized fragments and detecting these fragments using their mass-to-charge ratio. These two components, used together, allow

9792-439: The target molecule, most often with a proton exchange. This produces the species in relatively high amounts. In negative chemical ionization (NCI) the reagent gas decreases the impact of the free electrons on the target analyte. This decreased energy typically leaves the fragment in great supply. A mass spectrometer is typically utilized in one of two ways: full scan or selective ion monitoring (SIM). The typical GC–MS instrument

9894-412: The temperature of which is precisely controlled electronically. (When discussing the "temperature of the column," an analyst is technically referring to the temperature of the column oven. The distinction, however, is not important and will not subsequently be made in this article.) The rate at which a sample passes through the column is directly proportional to the temperature of the column. The higher

9996-495: The time required in traditional laboratory systems. Additionally, the systems are smaller, and more mobile, including units that are mounted in mobile analytical laboratories (MAL), such as those used by the United States Marine Corps Chemical and Biological Incident Response Force MAL and other similar laboratories, and systems that are hand-carried by two-person teams or individuals, much ado to

10098-498: The type of MS/MS analysis being performed. Types of analysis include product ion scan, precursor ion scan, selected reaction monitoring (SRM) (sometimes referred to as multiple reaction monitoring (MRM)) and neutral loss scan. For example: When Q1 is in static mode (looking at one mass only as in SIM), and Q3 is in scanning mode, one obtains a so-called product ion spectrum (also called "daughter spectrum"). From this spectrum, one can select

10200-456: The use of C as the labeling and the measurement of C- C ratios with an isotope ratio mass spectrometer (IRMS); an MS with a detector designed to measure a few select ions and return values as ratios. Gas chromatograph Gas chromatography ( GC ) is a common type of chromatography used in analytical chemistry for separating and analyzing compounds that can be vaporized without decomposition . Typical uses of GC include testing

10302-529: Was improved by the invention of capillary column, in which the stationary phase is coated on the inner wall of the capillary. The autosampler provides the means to introduce a sample automatically into the inlets. Manual insertion of the sample is possible but is no longer common. Automatic insertion provides better reproducibility and time-optimization. Different kinds of autosamplers exist. Autosamplers can be classified in relation to sample capacity (auto-injectors vs. autosamplers, where auto-injectors can work

10404-456: Was shown to lead to promising results in the identification of VOCs from raw GC–MS data. Another method of analysis measures the peaks in relation to one another. In this method, the tallest peak is assigned 100% of the value, and the other peaks being assigned proportionate values. All values above 3% are assigned. The total mass of the unknown compound is normally indicated by the parent peak. The value of this parent peak can be used to fit with