Misplaced Pages

Mach–Zehnder interferometer

Article snapshot taken from Wikipedia with creative commons attribution-sharealike license. Give it a read and then ask your questions in the chat. We can research this topic together.

The Mach–Zehnder interferometer is a device used to determine the relative phase shift variations between two collimated beams derived by splitting light from a single source. The interferometer has been used, among other things, to measure phase shifts between the two beams caused by a sample or a change in length of one of the paths. The apparatus is named after the physicists Ludwig Mach (the son of Ernst Mach ) and Ludwig Zehnder ; Zehnder's proposal in an 1891 article was refined by Mach in an 1892 article. Mach–Zehnder interferometry with electrons as well as with light has been demonstrated. The versatility of the Mach–Zehnder configuration has led to its being used in a range of research topics efforts especially in fundamental quantum mechanics.

#157842

68-460: The Mach–Zehnder check interferometer is a highly configurable instrument. In contrast to the well-known Michelson interferometer , each of the well-separated light paths is traversed only once. If the source has a low coherence length then great care must be taken to equalize the two optical paths. White light in particular requires the optical paths to be simultaneously equalized over all wavelengths , or no fringes will be visible (unless

136-473: A dielectric coating and must be modified if a metallic coating is used or when different polarizations are taken into account. Also, in real interferometers, the thicknesses of the beamsplitters may differ, and the path lengths are not necessarily equal. Regardless, in the absence of absorption, conservation of energy guarantees that the two paths must differ by a half-wavelength phase shift. Also beamsplitters that are not 50/50 are frequently employed to improve

204-414: A mirror . The two beams then pass a second half-silvered mirror and enter two detectors. The Fresnel equations for reflection and transmission of a wave at a dielectric imply that there is a phase change for a reflection, when a wave propagating in a lower- refractive index medium reflects from a higher-refractive index medium, but not in the opposite case. A 180° phase shift occurs upon reflection from

272-420: A (1 × wavelength + 2 k ) phase shift due to two front-surface reflections, one rear-surface reflection. Therefore, when there is no sample, only detector 1 receives light. If a sample is placed in the path of the sample beam, the intensities of the beams entering the two detectors will change, allowing the calculation of the phase shift caused by the sample. We can model a photon going through

340-414: A Fabry–Pérot system. Compared with Lyot filters, which use birefringent elements, Michelson interferometers have a relatively low temperature sensitivity. On the negative side, Michelson interferometers have a relatively restricted wavelength range, and require use of prefilters which restrict transmittance. The reliability of Michelson interferometers has tended to favor their use in space applications, while

408-492: A collimator. Michelson (1918) criticized the Twyman–Green configuration as being unsuitable for the testing of large optical components, since the available light sources had limited coherence length . Michelson pointed out that constraints on geometry forced by the limited coherence length required the use of a reference mirror of equal size to the test mirror, making the Twyman–Green impractical for many purposes. Decades later,

476-789: A genuine quantum superposition of the two paths. The Mach–Zehnder interferometer's relatively large and freely accessible working space, and its flexibility in locating the fringes has made it the interferometer of choice for visualizing flow in wind tunnels and for flow visualization studies in general. It is frequently used in the fields of aerodynamics, plasma physics and heat transfer to measure pressure, density, and temperature changes in gases. Mach–Zehnder interferometers are used in electro-optic modulators , electronic devices used in various fiber-optic communication applications. Mach–Zehnder modulators are incorporated in monolithic integrated circuits and offer well-behaved, high-bandwidth electro-optic amplitude and phase responses over

544-427: A glass plate. At detector 2, in the absence of a sample, the sample beam and reference beam will arrive with a phase difference of half a wavelength, yielding complete destructive interference. The RB arriving at detector 2 will have undergone a phase shift of (0.5 × wavelength + 2 k ) due to one front-surface reflection and two transmissions. The SB arriving at detector 2 will have undergone

612-452: A monochromatic filter is used to isolate a single wavelength). As seen in Fig. 1, a compensating cell made of the same type of glass as the test cell (so as to have equal optical dispersion ) would be placed in the path of the reference beam to match the test cell. Note also the precise orientation of the beam splitters . The reflecting surfaces of the beam splitters would be oriented so that

680-774: A multiple-gigahertz frequency range. Mach–Zehnder interferometers are also used to study one of the most counterintuitive predictions of quantum mechanics, the phenomenon known as quantum entanglement . The possibility to easily control the features of the light in the reference channel without disturbing the light in the object channel popularized the Mach–;Zehnder configuration in holographic interferometry . In particular, optical heterodyne detection with an off-axis, frequency-shifted reference beam ensures good experimental conditions for shot-noise limited holography with video-rate cameras, vibrometry, and laser Doppler imaging of blood flow. In optical telecommunications it

748-442: A polarizing Michelson Interferometer as a narrow band filter was first described by Evans who developed a birefringent photometer where the incoming light is split into two orthogonally polarized components by a polarizing beam splitter, sandwiched between two halves of a Michelson cube. This led to the first polarizing wide-field Michelson interferometer described by Title and Ramsey which was used for solar observations; and led to

SECTION 10

#1732780641158

816-424: A vacuum, which is 1. Specifically, its speed is: v  =  c / n , where c is the speed of light in vacuum , and n is the index of refraction. This causes a phase shift increase proportional to ( n  − 1) ×  length traveled . If k is the constant phase shift incurred by passing through a glass plate on which a mirror resides, a total of 2 k phase shift occurs when reflecting from

884-482: Is a Michelson interferometer. One interferometer arm is focused onto the tissue sample and scans the sample in an X-Y longitudinal raster pattern. The other interferometer arm is bounced off a reference mirror. Reflected light from the tissue sample is combined with reflected light from the reference. Because of the low coherence of the light source, interferometric signal is observed only over a limited depth of sample. X-Y scanning therefore records one thin optical slice of

952-893: Is a superposition of the "lower" path ψ l = ( 1 0 ) {\displaystyle \psi _{l}={\begin{pmatrix}1\\0\end{pmatrix}}} and the "upper" path ψ u = ( 0 1 ) {\displaystyle \psi _{u}={\begin{pmatrix}0\\1\end{pmatrix}}} , that is, ψ = α ψ l + β ψ u {\displaystyle \psi =\alpha \psi _{l}+\beta \psi _{u}} for complex α , β {\displaystyle \alpha ,\beta } such that | α | 2 + | β | 2 = 1 {\displaystyle |\alpha |^{2}+|\beta |^{2}=1} . Both beam splitters are modelled as

1020-439: Is collimated into a parallel beam. A convex spherical mirror is positioned so that its center of curvature coincides with the focus of the lens being tested. The emergent beam is recorded by an imaging system for analysis. The "LUPI" is a Twyman–Green interferometer that uses a coherent laser light source. The high coherence length of a laser allows unequal path lengths in the test and reference arms and permits economical use of

1088-536: Is eliminated by using extremely narrowband light from a laser. The extent of the fringes depends on the coherence length of the source. In Fig. 3b, the yellow sodium light used for the fringe illustration consists of a pair of closely spaced lines, D 1 and D 2 , implying that the interference pattern will blur after several hundred fringes. Single longitudinal mode lasers are highly coherent and can produce high contrast interference with differential pathlengths of millions or even billions of wavelengths. On

1156-469: Is employed in many scientific experiments and became well known for its use by Michelson and Edward Morley in the famous Michelson–Morley experiment (1887) in a configuration which would have detected the Earth's motion through the supposed luminiferous aether that most physicists at the time believed was the medium in which light waves propagated . The null result of that experiment essentially disproved

1224-425: Is generated by making measurements of the signal at many discrete positions of the moving mirror. A Fourier transform converts the interferogram into an actual spectrum. Fourier transform spectrometers can offer significant advantages over dispersive (i.e., grating and prism) spectrometers under certain conditions. (1) The Michelson interferometer's detector in effect monitors all wavelengths simultaneously throughout

1292-417: Is nonzero when optical path difference Δ L > ℓ coh {\displaystyle \Delta L>\ell _{\text{coh}}} exceeds coherence length of light beams. The nontrivial features of phase fluctuations in optical phase-conjugating mirror had been studied via Michelson interferometer with two independent PC-mirrors . The phase-conjugating Michelson interferometry

1360-446: Is perfect spatial alignment between the returning beams, then there will not be any such pattern but rather a constant intensity over the beam dependent on the differential pathlength; this is difficult, requiring very precise control of the beam paths. Fig. 2 shows use of a coherent (laser) source. Narrowband spectral light from a discharge or even white light can also be used, however to obtain significant interference contrast it

1428-422: Is required that the differential pathlength is reduced below the coherence length of the light source. That can be only micrometers for white light, as discussed below. If a lossless beamsplitter is employed, then one can show that optical energy is conserved . At every point on the interference pattern, the power that is not directed to the detector at E is rather present in a beam (not shown) returning in

SECTION 20

#1732780641158

1496-617: Is second-order correlation function, the interference curve in phase-conjugating interferometer has much longer period defined by frequency shift δ ω = Δ k c {\displaystyle \delta \omega =\Delta kc} of reflected beams: I ( Δ L ) ∼ [ 1 + [ γ ( Δ L ) + 0.25 ] cos ⁡ ( Δ k Δ L ) ] , {\displaystyle I(\Delta L)\sim [1+[\gamma (\Delta L)+0.25]\cos(\Delta k\Delta L)],} where visibility curve

1564-408: Is thereby more pronounced, and this can be used to construct an asymmetric optical interleaver. The reflection from phase-conjugating mirror of two light beams inverses their phase difference Δ φ {\displaystyle \Delta \varphi } to the opposite one − Δ φ {\displaystyle -\Delta \varphi } . For this reason

1632-644: Is used as an electro-optic modulator for phase and amplitude modulation of light. Optical computing researchers have proposed using Mach-Zehnder interferometer configurations in optical neural chips for greatly accelerating complex-valued neural network algorithms. The versatility of the Mach–Zehnder configuration has led to its being used in a wide range of fundamental research topics in quantum mechanics, including studies on counterfactual definiteness , quantum entanglement , quantum computation , quantum cryptography , quantum logic , Elitzur–Vaidman bomb tester ,

1700-425: Is used in a number of different applications. Fig. 5 illustrates the operation of a Fourier transform spectrometer, which is essentially a Michelson interferometer with one mirror movable. (A practical Fourier transform spectrometer would substitute corner cube reflectors for the flat mirrors of the conventional Michelson interferometer, but for simplicity, the illustration does not show this.) An interferogram

1768-417: Is used. In Fig. 2, we see that the fringes can be adjusted so that they are localized in any desired plane. In most cases, the fringes would be adjusted to lie in the same plane as the test object, so that fringes and test object can be photographed together. The collimated beam is split by a half-silvered mirror . The two resulting beams (the "sample beam" and the "reference beam") are each reflected by

1836-416: The quantum eraser experiment , the quantum Zeno effect , and neutron diffraction . Michelson interferometer The Michelson interferometer is a common configuration for optical interferometry and was invented by the 19/20th-century American physicist Albert Abraham Michelson . Using a beam splitter , a light source is split into two arms. Each of those light beams is reflected back toward

1904-475: The Mach–Zehnder interferometer to estimate the phase shift by estimating these probabilities. It is interesting to consider what would happen if the photon were definitely in either the "lower" or "upper" paths between the beam splitters. This can be accomplished by blocking one of the paths, or equivalently by removing the first beam splitter (and feeding the photon from the left or the bottom, as desired). In both cases there will no longer be interference between

1972-456: The Shepherd et al. technique of deriving winds and temperatures from emission rate measurements at sequential path differences, but the scanning system used by PAMI is much simpler than the moving mirror systems in that it has no internal moving parts, instead scanning with a polarizer external to the interferometer. The PAMI was demonstrated in an observation campaign where its performance

2040-575: The Sun's surface. When used as a tunable narrow band filter, Michelson interferometers exhibit a number of advantages and disadvantages when compared with competing technologies such as Fabry–Pérot interferometers or Lyot filters . Michelson interferometers have the largest field of view for a specified wavelength, and are relatively simple in operation, since tuning is via mechanical rotation of waveplates rather than via high voltage control of piezoelectric crystals or lithium niobate optical modulators as used in

2108-432: The Twyman–Green configuration in testing large optical components. A similar scheme has been used by Tajammal M in his PhD thesis (Manchester University UK, 1995) to balance two arms of an LDA system. This system used fibre optic direction coupler. Michelson interferometry is the leading method for the direct detection of gravitational waves . This involves detecting tiny strains in space itself, affecting two long arms of

Mach–Zehnder interferometer - Misplaced Pages Continue

2176-513: The Upper Atmosphere Research Satellite, UARS, (launched on September 12, 1991) measured the global wind and temperature patterns from 80 to 300 km by using the visible airglow emission from these altitudes as a target and employing optical Doppler interferometry to measure the small wavelength shifts of the narrow atomic and molecular airglow emission lines induced by the bulk velocity of the atmosphere carrying

2244-544: The addition of the Virgo interferometer in Europe, it became possible to calculate the direction from which the gravitational waves originate, using the tiny arrival-time differences between the three detectors. In 2020, India was constructing a fourth Michelson interferometer for gravitational wave detection. Fig. 7 illustrates use of a Michelson interferometer as a tunable narrow band filter to create dopplergrams of

2312-418: The advent of laser light sources answered Michelson's objections. The use of a figured reference mirror in one arm allows the Twyman–Green interferometer to be used for testing various forms of optical component, such as lenses or telescope mirrors. Fig. 6 illustrates a Twyman–Green interferometer set up to test a lens. A point source of monochromatic light is expanded by a diverging lens (not shown), then

2380-484: The beamsplitter which then combines their amplitudes using the superposition principle . The resulting interference pattern that is not directed back toward the source is typically directed to some type of photoelectric detector or camera . For different applications of the interferometer, the two light paths can be with different lengths or incorporate optical elements or even materials under test. The Michelson interferometer (among other interferometer configurations)

2448-476: The broad wavelength range and overall simplicity of Fabry–Pérot interferometers has favored their use in ground-based systems. Another application of the Michelson interferometer is in optical coherence tomography (OCT), a medical imaging technique using low-coherence interferometry to provide tomographic visualization of internal tissue microstructures. As seen in Fig. 8, the core of a typical OCT system

2516-552: The characterization of high-frequency circuits, and low-cost THz power generation. The Michelson Interferometer has played an important role in studies of the upper atmosphere , revealing temperatures and winds, employing both space-borne, and ground-based instruments, by measuring the Doppler widths and shifts in the spectra of airglow and aurora. For example, the Wind Imaging Interferometer, WINDII, on

2584-403: The context of large scale cosmic events (known as strong field tests ). A Michelson interferometer consists minimally of mirrors M 1 & M 2 and a beam splitter M (although a diffraction grating is also used ). In Fig 2, a source S emits light that hits the beam splitter (in this case, a plate beamsplitter) surface M at point C . M is partially reflective, so part of

2652-478: The coronal magnetic field for studies of variability in the extended solar atmosphere. HMI observations will help establish the relationships between the internal dynamics and magnetic activity in order to understand solar variability and its effects. In one example of the use of the MDI, Stanford scientists reported the detection of several sunspot regions in the deep interior of the Sun, 1–2 days before they appeared on

2720-642: The development of a refined instrument applied to measurements of oscillations in the Sun's atmosphere, employing a network of observatories around the Earth known as the Global Oscillations Network Group (GONG). The Polarizing Atmospheric Michelson Interferometer, PAMI, developed by Bird et al., and discussed in Spectral Imaging of the Atmosphere , combines the polarization tuning technique of Title and Ramsey with

2788-417: The direction of the source. As shown in Fig. 3a and 3b, the observer has a direct view of mirror M 1 seen through the beam splitter, and sees a reflected image M' 2 of mirror M 2 . The fringes can be interpreted as the result of interference between light coming from the two virtual images S' 1 and S' 2 of the original source S . The characteristics of the interference pattern depend on

Mach–Zehnder interferometer - Misplaced Pages Continue

2856-412: The emitting species. The instrument was an all-glass field-widened achromatically and thermally compensated phase-stepping Michelson interferometer, along with a bare CCD detector that imaged the airglow limb through the interferometer. A sequence of phase-stepped images was processed to derive the wind velocity for two orthogonal view directions, yielding the horizontal wind vector. The principle of using

2924-412: The entire measurement. When using a noisy detector, such as at infrared wavelengths, this offers an increase in signal-to-noise ratio while using only a single detector element; (2) the interferometer does not require a limited aperture as do grating or prism spectrometers, which require the incoming light to pass through a narrow slit in order to achieve high spectral resolution. This is an advantage when

2992-436: The existence of such an aether, leading eventually to the special theory of relativity and the revolution in physics at the beginning of the twentieth century. In 2015, another application of the Michelson interferometer, LIGO , made the first direct observation of gravitational waves . That observation confirmed an important prediction of general relativity , validating the theory's prediction of space-time distortion in

3060-407: The front of a mirror, since the medium behind the mirror (glass) has a higher refractive index than the medium the light is traveling in (air). No phase shift accompanies a rear-surface reflection, since the medium behind the mirror (air) has a lower refractive index than the medium the light is traveling in (glass). The speed of light is lower in media with an index of refraction greater than that of

3128-406: The incoming light is not of a single spatial mode. For more information, see Fellgett's advantage . The Twyman–Green interferometer is a variation of the Michelson interferometer used to test small optical components, invented and patented by Twyman and Green in 1916. The basic characteristics distinguishing it from the Michelson configuration are the use of a monochromatic point light source and

3196-434: The interference fringes will generally take the shape of conic sections (hyperbolas), but if M 1 and M' 2 overlap, the fringes near the axis will be straight, parallel, and equally spaced (fringes of equal thickness). If S is an extended source rather than a point source as illustrated, the fringes of Fig. 3a must be observed with a telescope set at infinity, while the fringes of Fig. 3b will be localized on

3264-595: The interference pattern in twin-beam interferometer changes drastically. Compared to conventional Michelson interference curve with period of half-wavelength λ / 2 {\displaystyle \lambda /2} : I ( Δ L ) ∼ [ 1 + γ ( Δ L ) cos ⁡ ( 2 k Δ L ) ] , {\displaystyle I(\Delta L)\sim [1+\gamma (\Delta L)\cos(2k\Delta L)],} where γ ( Δ L ) {\displaystyle \gamma (\Delta L)}

3332-496: The interferometer by assigning a probability amplitude to each of the two possible paths: the "lower" path which starts from the left, goes straight through both beam splitters, and ends at the top, and the "upper" path which starts from the bottom, goes straight through both beam splitters, and ends at the right. The quantum state describing the photon is therefore a vector ψ ∈ C 2 {\displaystyle \psi \in \mathbb {C} ^{2}} that

3400-534: The interferometer unequally, due to a strong passing gravitational wave. In 2015 the first detection of gravitational waves was accomplished using the two Michelson interferometers, each with 4 km arms, which comprise the Laser Interferometer Gravitational-Wave Observatory . This was the first experimental validation of gravitational waves, predicted by Albert Einstein 's General Theory of Relativity . With

3468-400: The interferometer's performance in certain types of measurement. In Fig. 3, in the absence of a sample, both the sample beam (SB) and the reference beam (RB) will arrive in phase at detector 1, yielding constructive interference . Both SB and RB will have undergone a phase shift of (1 × wavelength +  k ) due to two front-surface reflections and one transmission through

SECTION 50

#1732780641158

3536-457: The interferometer. Thereafter they switched to white (broadband) light, since using white light interferometry they could measure the point of absolute phase equalization (rather than phase modulo 2π), thus setting the two arms' pathlengths equal. More importantly, in a white light interferometer, any subsequent "fringe jump" (differential pathlength shift of one wavelength) would always be detected. The Michelson interferometer configuration

3604-411: The light is transmitted through to point B while some is reflected in the direction of A . Both beams recombine at point C' to produce an interference pattern incident on the detector at point E (or on the retina of a person's eye). If there is a slight angle between the two returning beams, for instance, then an imaging detector will record a sinusoidal fringe pattern as shown in Fig. 3b. If there

3672-452: The longitudinal and vector magnetic field over the entire visible disk thus extending the capabilities of its predecessor, the SOHO 's MDI instrument (See Fig. 9). HMI produces data to determine the interior sources and mechanisms of solar variability and how the physical processes inside the Sun are related to surface magnetic field and activity. It also produces data to enable estimates of

3740-476: The mirrors. White light has a tiny coherence length and is difficult to use in a Michelson (or Mach–Zehnder ) interferometer. Even a narrowband (or "quasi-monochromatic") spectral source requires careful attention to issues of chromatic dispersion when used to illuminate an interferometer. The two optical paths must be practically equal for all wavelengths present in the source. This requirement can be met if both light paths cross an equal thickness of glass of

3808-434: The nature of the light source and the precise orientation of the mirrors and beam splitter. In Fig. 3a, the optical elements are oriented so that S' 1 and S' 2 are in line with the observer, and the resulting interference pattern consists of circles centered on the normal to M 1 and M' 2 (fringes of equal inclination ). If, as in Fig. 3b, M 1 and M' 2 are tilted with respect to each other,

3876-492: The original wave as reflected by the other mirror. Because the phase change from the Gires–Tournois etalon is an almost step-like function of wavelength, the resulting interferometer has special characteristics. It has an application in fiber-optic communications as an optical interleaver . Both mirrors in a Michelson interferometer can be replaced with Gires–Tournois etalons. The step-like relation of phase to wavelength

3944-436: The other hand, using white (broadband) light, the central fringe is sharp, but away from the central fringe the fringes are colored and rapidly become indistinct to the eye. Early experimentalists attempting to detect the Earth's velocity relative to the supposed luminiferous aether , such as Michelson and Morley (1887) and Miller (1933), used quasi-monochromatic light only for initial alignment and coarse path equalization of

4012-436: The other path with a probability amplitude of i / 2 {\displaystyle i/{\sqrt {2}}} . The phase shifter on the upper arm is modelled as the unitary matrix P = ( 1 0 0 e i Δ Φ ) {\displaystyle P={\begin{pmatrix}1&0\\0&e^{i\Delta \Phi }\end{pmatrix}}} , which means that if

4080-421: The paths, and the probabilities are given by p ( u ) = p ( l ) = 1 / 2 {\displaystyle p(u)=p(l)=1/2} , independently of the phase Δ Φ {\displaystyle \Delta \Phi } . From this we can conclude that the photon does not take one path or another after the first beam splitter, but rather that it must be described by

4148-422: The photon is on the "upper" path it will gain a relative phase of Δ Φ {\displaystyle \Delta \Phi } , and it will stay unchanged if it is on the lower path. A photon that enters the interferometer from the left will then end up described by the state and the probabilities that it will be detected at the right or at the top are given respectively by One can therefore use

SECTION 60

#1732780641158

4216-404: The rear of a mirror. This is because light traveling toward the rear of a mirror will enter the glass plate, incurring k phase shift, and then reflect from the mirror with no additional phase shift, since only air is now behind the mirror, and travel again back through the glass plate, incurring an additional k phase shift. The rule about phase shifts applies to beamsplitters constructed with

4284-454: The same dispersion . In Fig. 4a, the horizontal beam crosses the beam splitter three times, while the vertical beam crosses the beam splitter once. To equalize the dispersion, a so-called compensating plate identical to the substrate of the beam splitter may be inserted into the path of the vertical beam. In Fig. 4b, we see using a cube beam splitter already equalizes the pathlengths in glass. The requirement for dispersion equalization

4352-514: The sample at a time. By performing multiple scans, moving the reference mirror between each scan, an entire three-dimensional image of the tissue can be reconstructed. Recent advances have striven to combine the nanometer phase retrieval of coherent interferometry with the ranging capability of low-coherence interferometry. Others applications include delay line interferometer which convert phase modulation into amplitude modulation in DWDM networks,

4420-420: The solar disc. The detection of sunspots in the solar interior may thus provide valuable warnings about upcoming surface magnetic activity which could be used to improve and extend the predictions of space weather forecasts. This is a Michelson interferometer in which the mirror in one arm is replaced with a Gires–Tournois etalon . The highly dispersed wave reflected by the Gires–Tournois etalon interferes with

4488-470: The test and reference beams pass through an equal amount of glass. In this orientation, the test and reference beams each experience two front-surface reflections, resulting in the same number of phase inversions. The result is that light travels through an equal optical path length in both the test and reference beams leading to constructive interference. Collimated sources result in a nonlocalized fringe pattern. Localized fringes result when an extended source

4556-451: The unitary matrix B = 1 2 ( 1 i i 1 ) {\displaystyle B={\frac {1}{\sqrt {2}}}{\begin{pmatrix}1&i\\i&1\end{pmatrix}}} , which means that when a photon meets the beam splitter it will either stay on the same path with a probability amplitude of 1 / 2 {\displaystyle 1/{\sqrt {2}}} , or be reflected to

4624-571: Was compared to a Fabry–Pérot spectrometer, and employed to measure E-region winds. More recently, the Helioseismic and Magnetic Imager ( HMI ), on the Solar Dynamics Observatory , employs two Michelson Interferometers with a polarizer and other tunable elements, to study solar variability and to characterize the Sun's interior along with the various components of magnetic activity. HMI takes high-resolution measurements of

#157842